Nellcor OxiMax Pulse Oximeters

nellcor-n-600x-tabletop-pulse-oximeter-a (1)

Models Offered:

  • N-595
  • N-600
  • N-600x

These pulse oximeters are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. They are intended for use with neonatal, pediatric, and adult patients during both no motion and motion conditions and for patients who are well or poorly perfused, in hospitals, hospital-type facilities, intra-hospital transport, and home environments.

Specifications:

  • Measurement Range:
    • SpO2 – 1% to 100%
    • Pulse Rate – 20 to 250 bpm
    • Perfusion Range – 0.03% to 20%
  • Accuracy:
    • Saturation
      • Adult – 70 to 100% ±2 digits
      • Adult and Neonate Low Sat – 60 to 80% ±3 digits
      • Neonate – 70 to 100% ±2 digits
      • Low Perfusion – 70 to 100% ±2 digits
      • Adult and Neonate with Motion – 70 to 100% ±3 digits
    • Pulse Rate
      • Adult and Neonate – 20 to 250 bpm ±3 digits
      • Low Perfusion – 20 to 250 bpm ±3 digits
      • Adult and Neonate with Motion – 20 to 250 bpm ±5 digits
  • Display Update Interval: 1 second
  • Power Requirements: Rated at 100 to 120 volts AC (nominal 120 VAC) or 220 to 240 volts AC (nominal 230 VAC), 20 volt/amps to be compliant with IEC 60601 sub-clause 10.2.2.
  • Battery:
    • Fully Charged – 7 hours running time
    • Lead Acid Battery
    • 6 Volts DC

Please Call for Pricing: (314) 994-9820