Alaris 8100 Pump Module

IV010512-PR_media_01

The 8100 Pump Module is a large volume pump designed to be used alongside the 8015 PC Unit. This model is popular among facilities that utilize infusion pumps for the delivery of fluids, medications, blood, and blood products using continuous or intermittent delivery through clinically acceptable routes of administration; such as, intravenous (IV), intra-arterial (IA), subcutaneous, epidural, enteral, or irrigation of fluid spaces. Designed for use on adults, pediatrics, and neonates.

Specifications

  • Software Version 9.17
  • Dimensions: 3.3″W x 8.9″H x 5.5″D
  • Weight: 2.5 lbs
  • Operating Temperature: 41 to 104 degrees F
  • Infusion Pressure Max: 654 mmhg
  • KVO Rate: Default 1 mL/h
  • Rate Accuracy: ±5% at rates between 1 and 999 mL/h and ±5.5% at rates <1 mL/h under following conditions:
    • Infusion Rate Range: 0.1 to 999 mL/h
    • Ambient Temperature: 68 ±4°F (20 ±2°C)
    • Source Container Height: 20 inches above the top of the Pump Module
    • Test Solution: Distilled Water
    • Distal Back pressure: 0 mmHg (0 kPa)
    • Needle: 18 gauge
    • Administration Set Model 2210

Please Call for Pricing: (314) 994-9820