Triple Channel Large Volume Infusion Pump

Product Specifications
Standards - UL 544CSA C22.0, No 125

Case Material - Impact resistant polycarbonate/ABS alloy

Dimensions:

     Height: 7.875 inches (20.00 centimeters)

     Width: 6 inches 815.24 centimeters)

     Depth: 2.10 inches (5.33 centimeters)

Weight - Approximately 5.1 lbs. (2.32 kg) includes Pole Clamp

Air-In-Line (Default) - 500 µl except for Neonatal which is 50µl

Occlusion Pressure (Default) - 15 psi except for Controller pressure device which is 3 ft H20

Operating Temperature - 50- 104º Fahrenheit (10º - 40º Celsius)

Storage Temperature - <95º Fahrenheit (<35º Celsius) for optimum battery life

Maximum Storage Temperature - 131º Fahrenheit (55º Celsius) 

Rate Range - 0.1 - 999 milliliter per hour (each channel)

Volume Range - 0.1 - 999 milliliter (each channel)

KVO Rate Range - 0.1 - 20.0 milliliter per hour

System Accuracy - ±5%

Power Consumption - 6 watts AC power. Use only IVAC MedSystem III AC Adapter, Model 1550

Batteries -  

     Main - Rechargeable NiCad Battery Pack Memory

     Back-up - Nonrechargeable Lithium

Battery charge - A fully charged battery has a minimum of 6 hours running time with all channels running at 125 milliliter per hour and backlighting usage of 2 minutes per hour.The main battery retains 80% of its capacity after 500 charging Cycles, and retains 90% of its capacity after 3 months of Continuous AC charging.

AC Adapter & Cord Length - Model 1550, 8.5 VDC @ 750 mA & 8.5 feet (2.59 meters)

AC Adapter Connector - 4 pin locking connector is standard on Model 2863. Detachable Connector is standard on Model 2860

Fuses - 3 amp fast-blow internal

Ground Continuity - Maximum 0.1 ohm

Leakage Current - Maximum 100 microamps